Keysight N777-C Series Tunable Laser Family User's Guide
3
You can generate input triggers in any of the following ways:
•
applying a trigger to the Input Trigger Connector on the rear panel
of your instrument,
•
setting Trigger Configuration to Loopback so that an output trigger
automatically generates an input trigger, or
•
using the :TRIGger1 command, see your N777-C Programming
Guide.
How to Use Output Triggering
You can configure your Tunable Laser instrument to output a trigger when
the instrument performs certain tasks.
To set your instrument's Output Trigger Configuration:
1
From the Control Instrument window, click Trigger VIEW tab.
2
Use the menu to select the Output Trigger. The following output
trigger options are available:
•
Disabled, the output trigger mode is disabled.
•
Step Finished, a trigger is output after every step of a sweep
finishes.
•
Sweep Finished, a trigger is output after a sweep cycle finishes.
•
Sweep Started, a trigger is output after a sweep cycle starts.
If you choose Step Finished and a Continuous sweep, the wavelength
NOTE
interval between hardware triggers is set by the Step parameter,
although, the sweep is not stepped.
N777-C Instrument's Front Panel Display
5