8. Using a Mains Adapter

You can operate this device using the aponorm
mains adapter (DC 6 V, 600 mA).
Only use the aponorm
available as an original accessory appropriate for your
supply voltage.
Ensure that neither the mains adapter or the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket
the blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.

9. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
18
®
by microlife
®
by microlife mains adapter
Error
Description
«ERR 1»
Signal too
weak
«ERR 2»
Error signal
18
in
4
«ERR 3»
No pressure
in the cuff
17
«ERR 5»
Abnormal
result
«HI»
Pulse or cuff
pressure too
high
«LO»
Pulse too
low
Potential cause and remedy
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension.Repeat the measurement,
keeping your arm still.
An adequate pressure cannot be gene-
rated in the cuff. A leak may have
occurred. Check that the cuff is cor-
rectly connected and is not too loose.
Replace the batteries if necessary.
Repeat the measurement.
The measuring signals are inaccurate
and no result can therefore be dis-
played. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
The pressure in the cuff is too high
(over 300 mmHg) OR the pulse is too
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
The pulse is too low (less than 40
beats per minute). Repeat the measu-
rement.*
®