Instructions for use Upper arm B A S I S C O N T R O L The No. 1 blood pressure monitor in German pharmacies Arrhythmia Blood Memory cells User detection pressure scale Also suitable for: Children Kidney disease Diabetic Pregnant women...
Page 3
aponorm Basis Control ® Switches, housing and accessories Display On/off button Systolic value Display Diastolic value M button (Memory) Pulse indicator Time button Battery level indicator Cuff socket Blood pressure evaluation indicator Mains adapter connector Stored value Battery compartment Pulse rate measurement active Cuff Cuff check Cuff connector...
Page 4
aponorm Basis Control ® Designated use: Dear Customer, This oscillometric blood pressure monitor serves This instrument was developed in collaboration for non-invasive measurement of the blood pres- with physicians and clinical tests prove its mea- sure of persons aged 12 years or older. surement accuracy to be very high.* It is clinically validated for patients suffering from If you have any questions or problems or want to...
Table of contents Using the monitor for the first time Safety, care, accuracy test and disposal 14 · Inserting the batteries · Safety and protection · Setting the date and time · Monitor care · Selecting the correct cuff · Cleaning the cuff ·...
1. Using the monitor for the first time Selecting the correct cuff Microlife offers different cuff sizes. Select the cuff Inserting the batteries size to match the circumference of your upper arm After having unpacked the monitor, insert the batte- (measured by close fitting in the centre of the upper ries.
Page 7
2. Check list for performing long bar) must lie over the artery which runs reliable measurements down the inner side of the arm. · Support your arm so it is relaxed. P lease read the enclosed flyer on the “8 golden rules of taking the blood pressure”.
blood pressure again after a short while (approx. -D or -E. In this case re-position the cuff 1 hour). If the reading is still too high in the and repeat the measurement. For other error repeat measurement, please contact your doctor messages, please read chapter 7 on page 12.
4. Data memory ponds to the following ranges defined by internation- al guidelines (ESH, ESC, JSK) (values in mmHg). This monitor automatically stores the last 30 readings. Range Systo- Diasto- Recommenda- tion Viewing the stored values Blood Seek medical Press the M-button briefly, when the monitor is pressure ≥...
Clearing all values In the event of a battery change or power If you are sure that you want to permanently failure, the memory retains all values mea- remove all stored values, hold down the M sured; the date and time must, however, be button (the monitor must have been switched off reset.
If you do not receive an error message, but 6. Using a mains adapter if the results seem unusual to you, please You can operate this monitor using the Microlife check whether you observed all points of the mains adapter (DC 6 V, 600 mA).
Page 12
Error Designa- Potential cuase and Error Designa- Potential cuase and tion remedy tion remedy «ERR 1» Signal too The pulse signals on «ERR 3» Irregular It is not possible to build weak the cuff are too weak. cuff pres- up sufficient pressure in Re-position the cuff and sure the cuff.
instructions carefully and retain them for Error Designa- Potential cuase and future reference. tion remedy · This device may only be used for the purposes «LO» Pulse too The pulse rate is too described in these instructions. The manufac- low (less than 40 beats turer will not accept any liability for damage per minute).
· Do not use this monitor if you think it is · Deviations between measurements taken damaged or if you notice anything unusual. by your doctor or in the pharmacy and those taken at home are quite normal, as these ·...
Accuracy test that the product is free from defects in material and workmanship. Should, despite proper handling We recommend to have this monitor tested for of the product, a defect occur within the 5-year accuracy every 2 years or after severe mechanical warranty period, WEPA Apothekenbedarf GmbH impact (e.g.
proof is the date of the proof of purchase or of the commission the repair of the product with costs or warranty card with purchase date filled in by the to request the return of the unrepaired product. distributor. Further customer rights The warranty period is not extended by the fact WEPA Apothekenbedarf GmbH &...
Page 17
Measuring range: 20 - 280 mmHg – blood This monitor complies with the requirements of the pressure Medical Device Directive 93/42/EEC. 40 - 200 beats per minute - Technical alterations reserved. pulse rate Cuff pressure display range: 0 - 299 mmHg Measuring resolution: 1 mmHg Static accuracy: pressure within...
Page 18
aponorm Basis Control – Warranty card ® Name of customer Serial number of monitor (SN) Date of purchase Pharmacy stamp and signature aponorm ®...
Here you also fi nd the instructions for use in German, Turkish, French and Russian for download. Sold by: WEPA Apothekenbedarf Manufactured by: GmbH & Co KG Microlife AG Am Fichtenstrauch 6-10 D-56204 Hillscheid Espenstrasse 139 Assistive appliances No.: CH-9443 Widnau 21.28.01.2103...