Analog Board

Analog Board
Figure 7-7. CG1100 boards.
7-8 | Dycor CG1100 Oxygen Analyzer
Analog Board
Removing the analog board
The analog board has a total of eight cable/connectors attached to the
board. Please make note of the position of the cables before removing
and use the drawing in Figures 7-6 and 7-10 as a reference.
NOTE
Remove the 9-connector LED ribbon cable (J2) from the upper right-
hand corner of the analog board.
Remove the 2-position VSO valve connector (TB2) from the top
middle of the board.
Remove the 3-position Power connector (TB3) from the upper left-
hand top of the board.
TB3
TB3
TB6
TB6
TB4
TB4
TB5
TB5
J4
J4
J1
J1
PCB Bracket
PCB Bracket

MCU Board

MCU Board
TB2
TB2
J2
J2
LED Board
LED Board