Ametek Dycor CG1100 Series User Manual page 80

Table of Contents

Advertisement

Enable Alarm 1
Alarm 1 High Limit
Alarm 1 Low Limit
Enable Alarm 2
Alarm 2 High Limit
Alarm 2 Low Limit
Enable Alarm 3
Alarm 3 High Limit
Alarm 3 Low Limit
Enable Alarm 4
Alarm 4 High Limit
Alarm 4 Low Limit
Output Status
The following functions are used to emulate the cg1000-RTP.
Flow High Limit
Flow Low Limit
Alarm 2 PPM Limit
Alarm 2 % Limit
Alarm 2 select
Alarm Config
Alarm status
A-8 | Dycor CG1100 Oxygen Analyzer
0 = disabled, 1 = oxygen, 2 = flow, 3 = pressure, 4 = mois-
75
ture
2c
2D
0 = disabled, 1 = oxygen, 2 = flow, 3 = pressure, 4 = mois-
76
ture, 5 = cg1000
2E
2F
0 = disabled, 1 = oxygen, 2 = flow, 3 = pressure, 4 = mois-
06
ture
A3
A4
0 = disabled, 1 = oxygen, 2 = flow, 3 = pressure, 4 = mois-
07
ture
A5
A6
56
Bit
Output
0x0002
system alarm
0x0004
Alarm #1
0x0008
Alarm #2
0x0010
Zero valve
0x0020
span valve
0x0100
Alarm #3
0x0200
Alarm #4
High limit for flow alarm
7A
Low limit for flow alarm
7B
1F
6D
6B
0 = % alarm, 1 = PPM alarm
5E
Bit
Function
Alarm 1 on high/low flow – uses 7A and 7B
0x04
0x10
limits
Alarm 2 on high oxygen – uses 1F or 6D limits
0x20
Alarm 2 on low oxygen – uses 1F or 6D limits
61
Bit
Function
Alarm 1 high oxygen – this is not working in
0x0001
V1.1 – V1.3
0x0002
Alarm 1 low oxygen – this is not working in V1.1
0x0100
– V1.3
0x0200
Alarm 2 high flow
Alarm 2 low flow