CALIBRATING THE CG1100-RTP
NOTE

Calibration

NOTE
1.
2.
3.
4.
5.
6.
7.

Aborting a Calibration

Cancel a calibration at any time by clicking the button that you used to
select the type of calibration (Span & Zero, Span Only, Zero Only), or by
clicking Cancel at the bottom of the screen.
All oxygen alarms are disabled during calibrations.
The span gas is the high o
gas is the low o
concentration calibration gas.
2
set up your analyzer for a calibration gas procedure.
see chapter 3 - Installation.
using the Setup tab of the CG1100 configurator software, enter the
calibration gas values to match the span and zero calibration gases
you intend to use to calibrate your system.
set the 3-way valve on the back of the cg1100 analyzer to Cal.
set the 3-way valve that delivers the calibration gases to Span Gas.
click the appropriate gas calibration button on the Calibrate tab
(Span & Zero, Span Only, Zero Only) for dual or single gas calibration.
once the o
reading has stabilized on the Status display, click the
2
Calibrate tab. in the Calibration dialog box, click Calibrate Span or
Calibrate Zero to calibrate.
when calibration ends set the 3-way valve that delivers the calibra-
tion gases to the Off.
set the 3-way valve on the back of the cg1100 to Gas In.
concentration calibration gas. The zero
2
Calibration | 5-1