Table of Contents

Advertisement

4-6 | Dycor CG1100 Oxygen Analyzer

Setup Tab

Defines system parameter settings.
Figure 4-4. Setup tab for configurator software.
Carrier gas
specify the sample gas stream being analyzed in the list box.
Pressure
Enter the inlet pressure.
Define calibration settings.
Concentration
Enter span gas in percent or PPM. initiate span gas. check
o
reading on the Status tab until o
Span
2
Then click the Calibration tab and then the Calibrate Span
button .
Enter zero gas in percent or PPM. initiate zero gas. check o
Zero
reading on the Status tab until o
click on Calibration tab and then the Calibrate Zero button .
reading stabilizes.
2
reading stabilizes. Then
2
2

Advertisement

Table of Contents
loading

Table of Contents