Calibrate (C)
c0
Abort calibration or verify.
Start automatic calibration (requires auto cal. option)
c1
Start automatic verification (requires auto cal. option)
c2
c20
save the cell millivolt reading for the zero gas.
c10
save the cell millivolt reading for the span gas.
c31
Manually calibrate based on previously saved zero and span mil-
livolt readings.
c40
Manual verify using current reading.
Example:
Inject zero gas. Wait until reading stabilizes.
Master:
Slave:
A[CR]
Inject span gas. Wait until reading stabilizes.
Master:
Slave:
A[CR]
Calibrate.
Master:
Slave:
A[CR]
>02C20??[CR]
>02C10??[CR]
>02C31??[CR]
Serial Communications | A-5