Electrical Input Port Match - Agilent Technologies 83400 Series User Manual

Lightwave source and receiver modules
Table of Contents

Advertisement

Performance Tests
Tests for Sources

3. Electrical Input Port Match

1 Connect the equipment as shown in
Figure 4-3. Electrical input port match test setup
2 Turn on the equipment, and allow them to warm up for 1 hour.
3 Use the Agilent 8702D's "Guided Setup" feature to perform an electrical-to-
electrical response test. Use the following settings:
Start frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
Stop frequency (Agilent 83403C): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 GHz
Stop frequency (Agilent 83402C): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 GHz
Stimulus power: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dBm
IF bandwidth: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz
4 The guided setup routine will step you through an S11, 1-port, electrical
calibration.
5 Place the Agilent 8702D in single-sweep mode, and use the marker functions
to locate the peak of the response.
6 Compare the measurement with the specification listed in
"Specifications and Regulatory
4-6
Figure
4-3.
Information".
Chapter 3,

Advertisement

Table of Contents
loading

This manual is also suitable for:

83402c83403c83410c83411c83411d83412b

Table of Contents