Summary of Contents for Agilent Technologies 8495D

  • Page 1 Agilent Technologies 8495D/K Attenuators Operating and Service Manual Agilent Part Number: 08495-90027 Printed in USA Print Date: July 2002 Supersedes: April 2002...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies...
  • Page 3: Table Of Contents

    • “Description” on page 2 • “Specifications” on page 6 • “Installation” on page 8 • “Operating Instructions” on page 9 • “Operator’s Check” on page 10 • “Replaceable Parts” on page 12 8495D/K Attenuators Operating And Service Manual iii...
  • Page 4 Certification Agilent Technologies, Inc., certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extent allowed by the Institute’s calibration facility,...
  • Page 5 Service and Support Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. You can find a list of local service representatives on the Web at: http://www.agilent.com/find/assist Click on “Contact Us”...
  • Page 6 The CSA mark is a registered trademark of the Canadian Standards Association. 1SM1-A This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product (CISPER 11, Clause 4). vi 8495D/K Attenuators Operating And Service Manual...
  • Page 7 If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply. 8495D/K Attenuators Operating And Service Manual vii...
  • Page 8 8495D/K Attenuators Operating And Service Manual...
  • Page 9: Instrument Definition

    An attenuator manufactured after the printing of this manual may have a serial prefix that is not listed above. This unlisted serial prefix indicates that the attenuator is different from those documented in this manual. 8495D/K Attenuators Operating And Service Manual 1...
  • Page 10: Description

    Description Description The Agilent 8495D and 8495K are 50-ohm, dc to 26.5 GHz, 0-70 dB (in 10 dB steps), coaxial attenuators with APC 3.5 connectors. Each attenuator is made up of four attenuation sections connected in cascade. Each section consists of a precision thin-film attenuator card, a lossless thru-line, and a ganged pair of edge line transmission lines.
  • Page 11 TKP12-100-P-TS-100-AU. The connector plug and contacts without the cable can be ordered from a Viking distributor. Information on replaceable parts is available in Table 3 on page Solenoid Cable Refer also to Figure 2 on page Connectors 8495D/K Attenuators Operating And Service Manual 3...
  • Page 12 2. Pin 1 (red wire) is common for all coils. Pins 2, 3 and 4 are not used. 3. Pin 6 is common for all coils. Pins 1, 7, 8, 12, and 14 are not used. Figure 2 Solenoid Cable Pin Configuration 4 8495D/K Attenuators Operating And Service Manual...
  • Page 13 (8495K) 0370-1091 Knob (84950) 08496-20008 Base TKP12-100-P-TS-100-AU Connector Plug and Contacts only without cable (8495K) 1. Available from Viking Industries, Inc., Chatsworth, California 91311. Figure 3 Typical Solenoid Coil Driver Circuit 8495D/K Attenuators Operating And Service Manual 5...
  • Page 14: Specifications

    1 W average, 100 W peak with maximum pulse width of 10 microseconds Capability Solenoid Drive For Agilent 8495K only. Solenoids Drive Coil Voltage Switching Current 8495K 20 to 30 Vdc 125 mA at 24V (approx.) Switching speed Maximum 20 milliseconds including settling time 6 8495D/K Attenuators Operating And Service Manual...
  • Page 15 Containers and materials identical to those used in factory packaging are available through Agilent Technologies offices. If the instrument is being returned to Agilent Technologies for servicing, attach a tag indicating the type of service required, return address, model number, and full serial number.
  • Page 16: Installation

    (P1) with the jack (J1) on the attenuator, and then pushing the plug over Instructions the jack. The plug is removed by grasping the ribbed sides of the plug and squeezing them together while pulling back until the plug clears the jack. 8 8495D/K Attenuators Operating And Service Manual...
  • Page 17: Operating Instructions

    (Jl). By applying the proper voltage and grounds to the proper pins of Jl, the attenuator will either increase or decrease the amount of attenuation as selected. 8495D/K Attenuators Operating And Service Manual 9...
  • Page 18: Operator's Check

    Fine tune oscillator frequency to obtain maximum meter indication. 4. Set attenuator and SWR meter range switch as shown in Table 4 verify that SWR meter indicates within the limits shown. 10 8495D/K Attenuators Operating And Service Manual...
  • Page 19 Refer to the Agilent 11713A Attenuator/Switch Driver Operating and Service Manual for programming instructions. NOTE The attenuators have no internal adjustments and should not be opened. If defective, return the attenuator to the nearest Agilent Technologies office for repair. 8495D/K Attenuators Operating And Service Manual 11...
  • Page 20: Replaceable Parts

    *Available from distributors of Viking Industries, Inc., Chatsworth CA 91311 Troubleshooting Troubleshooting consists of performing the “Operator’s Check” on page If the instrument does not perform within limits, return the instrument to Agilent Technologies. 12 8495D/K Attenuators Operating And Service Manual...

This manual is also suitable for:

8495k