Getting Started
Entering Calibration Data into an Agilent 8702A/B
To install from an external disk drive
If you have an HP 9122-series dual-sided disk drive connected to your
Agilent 8702A/B, use this procedure to enter the source or receiver's calibra-
tion data.
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word RECEIVER for SOURCE
and the word RCVR for SRC.
1 Locate the LIF-formatted calibration disk provided with the source.
2 Place the disk in the disk drive.
3 On the Agilent 8702A/B, press:
PRESET
LOCAL, SYSTEM CONTROLLER
CAL, CAL KITS & STDS, SOURCE [COEFF], CAL STD: SRC DISK
LOAD SRC DISK, READ FILE TITLES, LOAD <filename>.
1-12