Tests for Receivers
1. Demodulation Frequency Response
This performance test requires the use of a special lightwave reference
receiver that is calibrated using a dual-hetrodyne YAG system. Since these ref-
erence receivers are not commercially available, your instrument must be
returned to the factory to verify this specification.
2. RF Port Match
1 Connect the equipment as shown in
Figure 4-4. RF port match test setup
2 Turn on the equipment, and allow them to warm up for 1 hour.
Figure
4-4.
Performance Tests
Tests for Receivers