Disinfect Therapy Device - Weinmann homecare SOMNOvent auto-S Description Of Device And Instructions For Use

Automatic bilevel s/st therapy device with autotrilevel principle
Table of Contents

Advertisement

6.6 Disinfect therapy device

If required, e.g. following infectious diseases or unusual contamination, you can also dis-
infect the housing, the power cord and the breathing tube. Follow the instructions for use
for the disinfectant used. It is recommended that suitable gloves (e.g. household or dispos-
able gloves) are used for disinfecting.
Therapy device
We recommend terralin
Warning! Risk of electric shock!
Before disinfecting, it is essential to disconnect the power cord from the power sup-
ply connection and to remove the plug from the socket.
Ensure that no liquids penetrate the therapy device. Never immerse the therapy de-
vice in disinfectants or other liquids, otherwise this may lead to damage to the ther-
apy device and thus to a hazard to users and patients.
Caution!
Ensure that no liquids penetrate the therapy device via the pressure measurement
connection, otherwise there may be malfunctions.
The housing and the power cord of the therapy device are cleaned simply by wiping with
disinfectant.
Breathing tube
We recommend gigasept
For disinfecting, proceed exactly as for cleaning.
Caution!
The breathing tube may not be heated to over 70 °C. Sterilization is not permitted.
Accessories
For disinfecting/sterilizing the exhalation system, the breathing mask and the
SOMNOclick 300 respiratory air humidifier, see the section entitled "Disinfect and sterilize"
in the corresponding instructions for use.
®
protect as a disinfectant for disinfecting by wiping.
®
FF for disinfecting by immersion.