glucose oxidase; peroxi-
Glucose
dase; potassium iodide;
(GLU)
buff er; non-reacti ve
ingredients
bis (3',3"-diio-
do-4',4"-dihy-
droxy-5',5"-dinitro-
Albumin
phenyl)-3,4,5,6-tetrab-
(ALB)
romosulfonephthalein;
buff er; non-reacti ve
ingredients
copper acetate; diiso-
propylbenzene dihydro-
Creati nine
peroxide; 3,3',5,5'-tetra-
(CRE)
methylbenzidine; buff er;
non-reacti ve ingredients
Ascorbic
2,6-dichlorophenolind-
Acid
ophenol; buff er and
(ASC)
non-reacti ve ingredients
o-Cresolphthalein
Calcium
Complexon; buff er and
(CA)
non-reacti ve ingredients
Note: For the parameter arrangement and combinati on of diff erent Urine
Reagent Strips, please refer to the product informati on on the Reacti f
Urinalysis Reagent Strips kit box or canister label.
Ensure that the type of strip selected corresponds with the strip being used.
If not, it will be detected and display an error.
Use only Reacti f Urinalysis Reagent Strips with the Reacti f Touch Urine
Analyzer for accurate results.
Reacti f Touch
70
nal von minden GmbH · Manual
Detects glucose as low
as 50-100 mg/dL (2.5-5
mmol/L).
Detects albumin as low
as 20-30 mg/L in clinical
urine.
Detects urine creati nine
between 10 mg/dL (0.9
mmol/L) and 300 mg/dL
(26.5 mmol/L)
Detects ascorbic acid as
low as 5-10 mg/dL (0.28-
0.56 mmol/L).
Detects urine calcium
between 4mg/dL (1.0
mmol/L) and 40mg/dL (10
mmol/L)
Detects glucose as low as 80-100 mg/
dL (4-5 mmol/L).
Detects albumin as low as 20-30
mg/L in clinical urine.
Detects urine creati nine between 10
mg/dL (0.9 mmol/L) and 300 mg/dL
(26.5 mmol/L)
Detects ascorbic acid as low as 8-10
mg/dL (0.45-0.56 mmol/L).
Detects urine calcium between 4mg/
dL (1.0 mmol/L) and 40mg/dL (10
mmol/L)