List of Errors and Advisory Messages
No.
Descripti on
Low batt ery power
1
Failure of test table
2
3
No printer paper
Failure of automati c
4
calibrati on
5
Incorrect strip type
6
Missing strip
7
Misplaced strip
Inverted strip positi -
8
oned on the test table
9
Tilted strip
10
Dry strip
Failure to fi nd end
11
of strip
Reacti f Touch
66
nal von minden GmbH · Manual
Acti on
Replace the batt eries. See Secti on 10, Changing Batt eries.
1.
Make sure that the test table is in place. Move the test table in or out
of the instrument slightly to repositi on the test table.
2.
If the error remains, with the instrument powered on, unplug the pow-
er cord from rear of instrument and plug it back in. Turn the instrument
on by pressing the gray power butt on.
3.
If the error remains with the test table in place, contact your local
Customer Service and Support Center.
Replace the printer paper. See Secti on 3, Loading the Printer Paper or Label
Roll.
Clean the calibrati on strip. If the error remains aft er cleaning, contact your
local Customer Service and Support Center.
Ensure that the strip type selected in the Instrument Setup matches with
the strip being used. If the correct type of strip is being used, check the
instrument operati on by running another test using: a) a yellow and clear
specimen, or b) Quality control.
Repeat the test ensuring that the strip is positi oned on the test table.
Repeat the test ensuring that the strip is correctly positi oned on the test
table. If the error remains, examine the test table insert to ensure the small,
white line located near the ti p of the strip (on strip side of insert) is present
and not damaged. If this line is damaged or missing, contact your local Custo-
mer Service and Support Center
Repeat the test ensuring that the strip is correctly positi oned on the test
table.
Repeat the test ensuring that the strip is correctly positi oned on the test
table.
Repeat the test ensuring that the strip has been immersed in specimen.
Repeat the test ensuring that the strip is correctly positi oned on the test
table.