Note:  The user can input up to 200 specimens' informati on for the work list.
If the analyzer has the Strip Lockout functi on, the user needs to enter the canister
code for all strip canisters that are going to be used before testi ng (For further
instructi ons on how to enter the canister code, refer to Secti on 6 Specimen Test:
Entering the Canister Code). The quanti ty of specimen informati on that can be
inputt ed should be no more than the quanti ty of strips entered. For example, if 2
canister codes of 50 strips per canister are scanned, then the user will only be able
to input up to 100 specimens' informati on.
Note:  The functi on of identi fying urine color will not work in Conti nuous Test Mode. If
the color selecti on is set to Determined by Analyzer and Conti nuous Test Mode
is being used, the displayed and/or printed result for urine color will be Not
Entered.
Press
to enter the Test Preparati on screen.
The strip testi ng process is identi cal to a Quick Strip Test. Press
Testi ng screen.
To stop the test, press
screen.
Note: For further instructi ons regarding the procedures required for running the
Reacti f Urinalysis Reagent Strip test, refer to Secti on 6 Specimen Test: Strip
Test- Quick Test Mode.
Note: When the Work List is ON, test setti ngs should not be changed. If the test
setti ngs must be changed, exit the Work List fi rst.
on the Test Preparati on screen to return to the Work List
to show the
25