Cauti on: Any soluti ons other than those listed above may damage the test table
and insert.
2.
Place the insert and/or test table into the soluti on, making sure the white calibrati -
on bar on the test table remains above the liquid level.
Cauti on: Be sure the soluti on does not come in contact with the white calibrati on bar.
Do not cover the container while the test table is soaking. .
3.
Soak the table and insert for a minimum of 2 minutes and maximum of 10 minutes.
Do not soak longer than 10 minutes.
4.
Rinse the test table and insert thoroughly with water.
5.
Dry with a soft cloth and replace test table and table insert in the analyzer (as de-
scribed on the previous page).
White Calibration Bar
Cleaning the White Calibrati on Bar
For the Reacti f Touch Analyzer to perform as intended and provide reliable test results,
it is recommended to periodically check the white calibrati on bar on the test table. For
proper use, the white calibrati on bar should not be dirty or discolored.
1.
Remove the insert from the test table.
2.
Remove the test table by pulling it slowly out of the analyzer.
3.
Check the white calibrati on bar on the test table for dirt or discolorati on.