Publication No. 980855 Rev. A
Setting the Message
Termination
Characters
EADS North America Test and Services
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Menu preferences are not normally retained after the power is
removed from the instrument. To make (a) menu preference(s)
permanent, make the desired change(s) to settings first and
then proceed to the section Storing the Menu Preferences to
make the changes permanent. Following this procedure will
put the settings into non-volatile memory and allow them to
be restored during instrument power-up.
Each message to or from the 1256 must end with an end-of-line
(EOL) character known as a terminator. You may select the EOL
terminator to be any one of the following:
Carriage return (CR) only
Line feed (LF) only (for non-terminal-based applications,
such as controlling via LabVIEW programs)
Carriage return and line feed (CR-LF)
The factory default is line feed only.
To select the EOL terminator, proceed as follows:
1. Press the top key. Rotate the knob until the top line reads:
>-----SETTINGS-----<
2. Select line 2 (next to top line) by pressing the second key.
This selects line 2 so that you may choose the type of setting
to work with:
-----SETTINGS-----
>
Type: Display
Option: Format
Status: Decimal
3. Rotate the knob until line 2 reads as shown below (lines 3
and 4 may differ):
-----SETTINGS-----
>
Type: RS-232
Option: Data Bits
Status: 8
4. Press the third key to select line 3 of the display. Rotate the
knob until line 3 appears as shown below:
-----SETTINGS-----
Type: RS-232
> Option: EOL Mode <
Status: CR Only
1256 User Manual
NOTE:
<
<
GPIB and RS-232 Interfaces 4-13