STATUS BYTE
Register
EADS North America Test and Services
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
For example, if the PON and QYE bits of the Standard Event
Status Register are set, but the Standard Event Status Enable
Register value is 0, then the ESB of the Status Byte Register will
not be set. If either, or both, bits 7 and bit 2 of the Standard Event
Status Enable Register are set, then the ESB bit of the Status Byte
Register will be set.
Another way of viewing the Standard Event Status Enable Register
is that it selects which conditions reflected in the Standard Event
Status Register are enabled to set the ESB bit of the status byte.
The Status Byte Register is similar to the Standard Event Status
Register. Each bit of this register reflects the true or false condition
of the corresponding bit. These bits reflect the PRESENT value of
the condition, whereas the Standard Event Status Register bits are
latched. That is, once a bit in the Standard Event Status Register
is set, it remains set until a *CLS command is executed or an
*ESR? Query is executed. However, the bits of the Status Byte
Register change states as the corresponding condition becomes
true or false. These bits are NOT latched.
Also, the Status Byte is not cleared by reading the register. Each
bit of the Status Byte remains set until the condition indicated by
the bit is no longer present.
The following bits are assigned in the Status Byte Register. All
other bits are not used and will return "0" when read with the
*STB? query:
OSE
Operation Status Event
Bit 7, bit weight = 128 decimal = 80 hexadecimal
This bit is set when any of the bits of the Operation
Status Event Register are set. (This bit is NOT
shown on the diagram. For a description of the
Operation Status Event Register, consult the "SCPI
Status Registers" section of this chapter.
MSS
Master Summary Status.
Bit 6, bit weight = 64 decimal = 40 hexadecimal
This bit is set when one or more of the "enabled"
bits of the Status Byte are set. In other words:
1256 User Manual
SCPI Command Basics 5-11