SCPI Command Basics 5-22
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The Operation Status Event Register latches the status
information. Once a bit is set in the Operation Status Event
Register, it remains set until the bit is cleared by reading the
register with the STATUS:OPERATION:EVENT? query, or by
sending the *CLS command.
When any of the bits of the Operation Status Event Register are
set, bit 7 of the Status Byte Register will be set.
The Questionable Status Register is not used by the 1256. When
the Questionable Status Condition Register or Questionable Status
Event Register are read, they will return a value of 0. The
Questionable Status Event Register may be programmed and
queried, but will have no effect on the operation of the 1256.
The following SCPI command tree shows the syntax of the SCPI
STATUS commands:
:STATus
:OPERation
[:EVENt]?
:CONDition?
:ENABle
:QUEStionable
[:EVENt]?
:CONDition?
:ENABle
Examples of the commands are shown below. To enable the "Wait
for Trigger" and "Wait for Arm" bits the following command
achieves this:
STAT:OPER:ENABLE 96
To read the value of the enable register:
STAT:OPER:ENABLE?
To read the value of the event register (and clear the event
register):
STAT:OPER?
To read the value of the condition register:
STAT:OPER:COND?
These commands augment the IEEE-488.2 Common Commands
to provide additional status information.