PDATAOUT

Operation 3-8
PDATAOUT <module address>;<module address>.....
This command causes the specified switching modules to output
the present state of the relays contained in those modules. An
example character string output by the Option 01 response to a
"PDATAOUT 0-12" is given below.
000. Model 1260 Universal Switch Controller
000. OS Rev 1.2 1260
012 1260-40A 4 x 24 signal matrix
012. 0219
012.END
This example shows that a 1260 Controller with an operating
system revision level of 1.2 is present at module address 000.
There is also a 1260-40A module present at address 012, and row
2 of the switching matrix is connected to column 19.
The module address for PDATAOUT and PSETUP
commands may specify a range of module addresses
by using the syntax <start module address>-<end
module address>. The Option 01 will output data for
all modules between the starting and ending module
addresses that are installed in the 1260.
The switching module configuration is described in
the same format as is required by the OPEN and
CLOSE commands.
The address obtained for the 1260 Universal Switch
Controller is not the address of the slot 0 Resource
Manager or the logical address of the 1260-01. It is
equivalent to the module address of a switching
module, and is always 000.
The character string output by the Option 01 is made
up of multiple lines and contains carriage-return, line-
feed pairs. These may terminate the output string,
and so the PSETUP output string must be read
repeatedly until all lines of the output have been read
for all modules specified. The final line output by the
1260 Series User Manual