5.4 ENVIRONMENTAL SPECIFICATIONS

Environmental specifications are based on component margins, thermal verification testing and
current draw tests. Production unit performance is not verified over temperature.
PARAMETER
Operating Temperature
Temperature Monitor Range
Channel Operating Temperature
AC Power Supply
Power Consumption
Chassis
Channel (per)
12 or 20GHz Channel (per)
Warm-Up Time
1
Specifications are subject to change per the discretion of Holzworth Instrumentation, Inc
*NOTE: Amplitude accuracy may vary when operated outside of channel operating temperature
specified in section 6.3. Internal channel temperature should be monitored using the ":TEMP?"
command (see Appendix A). External cooling is recommended to regulate internal temperature
depending on ambient operating conditions and number of integrated channels.
DESCRIPTION
Operating Environment
Humidity
Altitude
Vibration
Storage (Non-Operating)
Temperature
Humidity
Altitude
Vibration
HS9000B Series Manual 1.04
MIN
TYPICAL
0 C
-40 C
+35 C
+40 C
*
100 V
AC
5 W
7 W
15 W
10 min
SPECIFICATION (by design)
RH 20% to 80% at wet bulb temp. <29C (non-condensing)
0 to 2,000m (0 to 6,561 feet)
0.21 G-rms maximum, 5Hz to 500Hz
-10C to + 60C
RH 20% to 80% at wet bulb temp. <40C (non-condensing)
0 to 4,572m (0 to 15,000 feet)
Altitude
0.5 G-rms maximum, 5Hz to 500Hz
www.holzworth.com
MAX
COMMENTS
+55 C
+85 C
Absolute, channel dedicated outputs
+45 C
Amplitude accuracy valid range
240 V
47 – 63Hz
AC
20 min
20 C (ambient temp. dependent)