  • Page 1 HS9000B Series Multi-Channel RF Synthesizers RoHS User Manual 1.04 Holzworth Instrumentation Inc. 2540 Frontier Ave., Suite 200 Boulder, CO 80301 USA www.holzworth.com...
  • Page 2 This page intentionally left blank.
  • Page 3: Table Of Contents

    2.1 CE CERTIFICATION ......................3 2.2 RoHS EXEMPTION ....................... 3 3.0 PRODUCT WARRANTY ......................3 4.0 CALIBRATION NOTICE ......................3 5.0 HS9000B SERIES CONFIGURATION GUIDE ................. 4 5.1 CONFIGURATION SUMMARY ....................4 5.2 HARDWARE CONFIGURATION ................... 5 5.1.1 Number of Channels ....................... 5 5.1.2 Loaded Channel Frequencies ..................
  • Page 4 9.2.2 Reference Output Summary ..................28 10.0 CONTACT INFORMATION ....................29 APPENDIX A: PROGRAMMING COMMANDS ................30 APPENDIX B: ETHERNET CONFIGURATION COMMANDS ............60 APPENDIX C: GPIB CONFIGURATION COMMANDS ..............64 www.holzworth.com HS9000B Series Manual 1.04 Page 2 of 65...
  • Page 5: Introduction

    Holzworth Instrumentation Inc. This User‟s Manual is a generic, quick reference guide for use with the Holzworth HS9000B Series Multi-Channel RF Synthesizer products. Refer to section 5 for specific configuration details with regards to the HS9000B Series hardware.
  • Page 6: Hs9000B Series Configuration Guide

    Specific attention is paid to phase coherency between the independely controllable channels. The HS9000B Series is a unique platform allowing the user to specify custom configurations for a COTS product. Units are loaded with anywhere from 1 to 8 channels (up to 6.7GHz), with the additional flexibility to specify each channel‟s frequency limits and performance options.
  • Page 7: Hardware Configuration

    Additional factory loaded options are also defined in the “Loaded Options” designator on the front panel. These options further customize the HS9000B Series to an application and are loaded at the factory when the unit is initially built. Accessories are external to the HS9000B platform and can be ordered separately.
  • Page 8: Part Number Example

    Option OPT-REFX replaces the standard reference module with a reference module that allows the user to utilize various external reference frequencies. Allowable external reference frequencies with OPT-REFX are any 100kHz increment within the range of 5MHz - 160MHz. www.holzworth.com HS9000B Series Manual 1.04 Page 6 of 65...
  • Page 9: Mechanical Configuration

    5.3 MECHANICAL CONFIGURATION The HS9000B Series comes in a 1U high, rack mountable chassis. The example shown is of a 6 channel unit (front panel configuration may vary). A universal rack mount bracket kit is an available accessory (Part No.: RACK-1U). Mechanical dimensions are listed in inches (and millimeters).
  • Page 10: Environmental Specifications

    20 min 20 C (ambient temp. dependent) Specifications are subject to change per the discretion of Holzworth Instrumentation, Inc *NOTE: Amplitude accuracy may vary when operated outside of channel operating temperature specified in section 6.3. Internal channel temperature should be monitored using the ":TEMP?"...
  • Page 11: Multi-Channel Installation

    6.0 MULTI-CHANNEL INSTALLATION Prior to use of an HS9000B Series Multi-Channel Synthesizer users will need to be setup with the basic hardware and software. 6.1 HARDWARE INSTALLATION Prior to initializing the synthesizer, connect the power cord to an active AC power supply. The instrument is shipped with the appropriate power cord for the final destination country/region.
  • Page 12: Hs9000B Synthesizer Communication

    COM port the instrument is using. The COM port associated with the USB connection to the HS9000B can be identified by using the application GUI as shown above or via the Windows Device Manager. www.holzworth.com HS9000B Series Manual 1.04 Page 10 of 65...
  • Page 13: Usb Communication Troubleshooting

    Click on "Hardware and Sound" STEP TWO Under "Devices and Printers," select Device Manager STEP THREE Under Ports (COM & LPT) locate COM port associated with the HA7062C (identified as "USB Serial Port") www.holzworth.com HS9000B Series Manual 1.04 Page 11 of 65...
  • Page 14: Rs-232 Hardware Specifications

    5. Carriage Return: Carriage return (ASCII Code 13) 6. Pinout: Label Label Label TX (Response Output) RX (Instruction Input) 7.1.3 GPIB Communication The Holzworth HS9000B synthesizers are GPIB capable. GPIB configuration commands are listed in Appendix C. www.holzworth.com HS9000B Series Manual 1.04 Page 12 of 65...
  • Page 15: Ethernet Communication

    When the HS9000B is connected to a network with a DHCP server, the network settings will be auto-assigned per the servers configuration. If the HS9000B is connected directly to a PC or to a network with no DHCP server the instrument's default IP address will be: 169.254.117.11 www.holzworth.com HS9000B Series Manual 1.04 Page 13 of 65...
  • Page 16: Assigning A Static Ip Address

    5. Power cycle the instrument when prompted. When the instrument fully powers back on (5-10 second power up) it will come up with the static IP settings and can be connected to the LAN. www.holzworth.com HS9000B Series Manual 1.04 Page 14 of 65...
  • Page 17: Troubleshooting Ethernet Connections

    5. Begin by querying with the :IP:STATUS? command. Change status and/or re-configure the static network settings as necessary. 6. Power cycle the HS9000B if prompted. Any status change from DHCP to Static or vice versa will require a power cycle. www.holzworth.com HS9000B Series Manual 1.04 Page 15 of 65...
  • Page 18: Miscellaneous Ethernet Troubleshooting

    6. Make a direct Ethernet connection from the PC to the synthesizer bypassing any routers or network switches. 7. Right click the network adapter the synthesizer is connected to and click Properties. www.holzworth.com HS9000B Series Manual 1.04 Page 16 of 65...
  • Page 19 The PC IP address will default to 169.254.xxx.xxx and the subnet address will default to 255.255.0.0. 7. Close and re-launch the application GUI. Attempt to establish a connection to the HS9000B. 8. For further assistance please contact Holzworth Support. www.holzworth.com HS9000B Series Manual 1.04 Page 17 of 65...
  • Page 20: Application Gui Operation

    Highlighted Field both left and right Up/Down Arrows used to increase/decrease the value of the Highlighted Field Number Keys used to directly enter value into active field www.holzworth.com HS9000B Series Manual 1.04 Page 18 of 65...
  • Page 21: Channel Enable Function

    In the case of operating a synthesizer equipped with less than 8 channels, only channels 1 though N will be displayed for control, the remaining channel slots will not be available. www.holzworth.com HS9000B Series Manual 1.04 Page 19 of 65...
  • Page 22: Reference Menu

    As previously noted the REFX reference frequency may be any 100kHz increment in the range of 5MHz to 160MHz www.holzworth.com HS9000B Series Manual 1.04 Page 20 of 65...
  • Page 23: Modulation Menu

    Radio buttons select channel Change view to Mod. menu Change view to modulation Turn on/off each type of for modulation, or apply to ALL to set AM, FM, PM, etc. type sub-menu's modulation www.holzworth.com HS9000B Series Manual 1.04 Page 21 of 65...
  • Page 24: Pulse Modulation

    On/Off Ratio 10MHz to 2GHz > 60dB 2GHz to 5GHz > 50dB 5GHz to 18GHz > 90dB Minimum Pulse Width 50 ns ALC Loop Deviation (ALC disabled) 1dB difference from ALC enabled www.holzworth.com HS9000B Series Manual 1.04 Page 22 of 65...
  • Page 25: Frequency Modulation(Channels Up To 6.7Ghz, External Mod. Stimulus)

    DC to 10 kHz (-3dB) DC Coupled + 1V: Maximum Amplitude ± 1V peak for indicated Sensitivity when using Ext. Input 0V: 50% of Maximum Depth Depth (into 50Ω) - 1V: Maximum Depth www.holzworth.com HS9000B Series Manual 1.04 Page 23 of 65...
  • Page 26: Sweep Mode

    TRIGGER SWEEP Point: Selecting the Point sweep mode will initiate each individual point step in the sweep bandwidth at each trigger signal. NOTE that once sweep mode is initiated, it will loop (restart) continuously until the routine is manually interrupted. www.holzworth.com HS9000B Series Manual 1.04 Page 24 of 65...
  • Page 27: List Mode (Channels Up To 6.7Ghz)

    NOTE 2: Narrow List Mode list does not allow for amplitude control. NOTE 3: The 5% frequency range limitation for Narrow List Mode is defined as: the center frequency (f ) ± 2.5%. www.holzworth.com HS9000B Series Manual 1.04 Page 25 of 65...
  • Page 28 SAVE TO FILE: The Save to File feature is used to save a .csv file of whatever list command line data is currently loaded in the List Transfer Window. www.holzworth.com HS9000B Series Manual 1.04 Page 26 of 65...
  • Page 29: Firmware Updates

    4. Do not turn off, unplug, or disconnect USB from the analyzer/PC while updates are in progress. When the progress bar reaches 100% and the updater window displays "Update complete" close the updater window. www.holzworth.com HS9000B Series Manual 1.04 Page 27 of 65...
  • Page 30: Rf Inputs/Outputs

    9.0 RF INPUTS/OUTPUTS The HS9000B Series Multi-Channel RF Synthesizers are CW work horses. They are designed to do an excellent job of providing highly stable, phase coherent signals with pure spectrums and highly accurate output power amplitude control. 9.1 RF OUTPUTS The RF Output ports are labeled and positioned sequentially from left to right on the front panel of the instrument.
  • Page 31: Contact Information

    Contact Holzworth directly for product support. A list of US Sales Representatives and non-US Distribution partners are listed on the Holzworth website. Holzworth Instrumentation Sales Support Phone: +1.303.325.3473 (option 1) Email: sales@holzworth.com Holzworth Instrumentation Technical Support Phone: +1.303.325.3473 (option 2) Email: support@holzworth.com www.HOLZWORTH.com www.holzworth.com HS9000B Series Manual 1.04 Page 29 of 65...
  • Page 32: Appendix A: Programming Commands

    APPENDIX A: PROGRAMMING COMMANDS The Holzworth Instrumentation HS9000B Series Synthesizers allow users to communication with the instrument over USB, Ethernet, RS-232, or GPIB using their own application software. The programming commands are ASCII commands, and an internal communications module will forward the commands to the appropriate synthesizer channel.
  • Page 33 Query if the communications bus is ready to receive additional commands Example :COMM:READY? Communications Bus Ready <OR> Communications Bus is Busy :ATTACH? Query the number of internal channels Example :ATTACH? :CH1:CH2 <OR> :CH1:CH2:CH3:CH4 <OR> etc www.holzworth.com HS9000B Series Manual 1.04 Page 31 of 65...
  • Page 34 *RCL Recall Saved State Example :CHn*RCL State Recalled *SAV Save Current State Example :CHn*SAV State Saved :IDN? Identify Example :CHn:IDN? Holzworth,HSM6001A,M1009-001,FW3.31,HS9002A-112 (Manufacturer, Device Name, Board #, Firmware version, Instrument Serial #) www.holzworth.com HS9000B Series Manual 1.04 Page 32 of 65...
  • Page 35 Read Temperature :TEMP? Query the temperature of the channel Example :CHn:TEMP? Temp = 40C www.holzworth.com HS9000B Series Manual 1.04 Page 33 of 65...
  • Page 36 :FREQ? Query Synthesizer RF Frequency Example :CHn:FREQ? 22.67 MHz :FREQ:MAX? Query Synthesizer Maximum RF Set Frequency Example :CHn:FREQ:MAX? 1.024 MHz :FREQ:MIN? Query Synthesizer Minimum RF Set Frequency Example :CHn:FREQ:MIN? 0.25 MHz www.holzworth.com HS9000B Series Manual 1.04 Page 34 of 65...
  • Page 37 Power Set :PWR? Query Synthesizer RF Power Example :CHn:PWR? 9.50 :PWR:MAX? Query Synthesizer Maximum RF Set Power Example :CHn:PWR:MAX? 10.00 dBm :PWR:MIN? Query Synthesizer Minimum RF Set Power Example :CHn:PWR:MIN? -100.00 dbm www.holzworth.com HS9000B Series Manual 1.04 Page 35 of 65...
  • Page 38 Phase Set :PHASE? Query Synthesizer RF Phase Offset Example :CHn:PHASE? 270.1 :PHASE:MAX? Query Synthesizer Maximum RF Phase Offset Example :CHn:PHASE:MAX? 359.9deg :PHASE:MIN? Query Synthesizer Minimum RF Phase Offset Example :CHn:PHASE:MIN? 0.0deg www.holzworth.com HS9000B Series Manual 1.04 Page 36 of 65...
  • Page 39 Set RF ON/OFF :PWR:RF:<value> Set Synthesizer RF ON/OFF <value> ON <or> OFF Example :CHn:PWR:RF:ON RF POWER ON :PWR:RF? Query Synthesizer RF ON/OFF Example :CHn:PWR:RF? ON <or> OFF www.holzworth.com HS9000B Series Manual 1.04 Page 37 of 65...
  • Page 40 NOTE: In the value field above, PULSE or PULSE:SRC:EXT enable external pulse modulation. :MOD:MODE? Query Modulation Mode Status Example :CHn:MOD:MODE? OFF <or> PULSE:EXT <or> PULSE:INT <or> PULSE:INT:TRIGGER <or> FM <or> AM <or> PM <or> SWEEP:FREQ <or> LOOKUP:NARROW <or> LOOKUP:WIDE www.holzworth.com HS9000B Series Manual 1.04 Page 38 of 65...
  • Page 41 Set Synthesizer FM Deviation <value> Synthesizer Dependent [suffix] Hz, kHz Example :CHn:MOD:FM:DEV:1.2kHz FM Deviation Set :MOD:FM:DEV? Query Synthesizer FM Deviation Example :CHn:MOD:FM:DEV? 0.500 kHz :MOD:FM:DEV:MAX? Query Synthesizer Maximum FM Deviation Example :CHn:MOD:FM:DEV:MAX? 100.000 kHz www.holzworth.com HS9000B Series Manual 1.04 Page 39 of 65...
  • Page 42 Set Synthesizer AM Depth <value> Synthesizer Dependent [suffix] [percent] Example :CHn:MOD:AM:DEPTH:15 percent AM Depth Set :MOD:AM:DEPTH? Query Synthesizer AM Depth Example :CHn:MOD:AM:DEPTH? 60 percent :MOD:AM:DEPTH:MAX? Query Synthesizer Maximum AM Depth Example :CHn:MOD:AM:DEPTH:MAX? 75 percent www.holzworth.com HS9000B Series Manual 1.04 Page 40 of 65...
  • Page 43 <value> Synthesizer Dependent [suffix] [deg] Example :CHn:MOD:PM:DEV: 45 deg PM Deviation Set :MOD:PM:DEV? Query Synthesizer PM Deviation Example :CHn:MOD:PM:DEV? 10 deg :MOD:PM:DEV:MAX? Query Synthesizer Maximum PM Deviation Example :CHn:MOD:PM:DEV:MAX? 180 deg www.holzworth.com HS9000B Series Manual 1.04 Page 41 of 65...
  • Page 44 <suffix> s, ms, [us] Example :CHn:MOD:PULSE:REP:45ms Pulse Rep Rate Set :MOD:PULSE:REP? Query Internal Pulse Repetition Rate Example :CHn:MOD:PULSE:REP? 45000.0 us :MOD:PULSE:REP:MAX? Query Maximum Internal Pulse Repetition Rate Example :CHn:MOD:PULSE:REP:MAX? 10000000.0 us www.holzworth.com HS9000B Series Manual 1.04 Page 42 of 65...
  • Page 45 <value> Synthesizer Dependent <suffix> s, ms, [us] Example :CHn:MOD:PULSE:WIDTH:45ms Pulse Width Set :MOD:PULSE:WIDTH? Query Internal Pulse Width Example :CHn:MOD:PULSE:WIDTH? 45000.0 us :MOD:PULSE:WIDTH:MAX? Query Maximum Internal Pulse Width Example :CHn:MOD:PULSE:WIDTH:MAX? 10000000.0 us www.holzworth.com HS9000B Series Manual 1.04 Page 43 of 65...
  • Page 46 Query Number of Output Pulses Example :CHn:MOD:PULSE:NUM? :MOD:PULSE:NUM:MAX? Query Maximum Number of Output Pulses Example :CHn:MOD:PULSE:NUM:MAX? 65535 The number of output pulses only applies when using Internal Pulse Modulation with a Trigger. www.holzworth.com HS9000B Series Manual 1.04 Page 44 of 65...
  • Page 47 * The maximum and minimum for the Sweep Start Frequency are the same as the corresponding values for the Set Frequency. Refer to the Set Frequency page for the maximum and minimum values. www.holzworth.com HS9000B Series Manual 1.04 Page 45 of 65...
  • Page 48 * The maximum and minimum for the Sweep Stop Frequency are the same as the corresponding values for the Set Frequency. Refer to the Set Frequency page for the maximum and minimum values. www.holzworth.com HS9000B Series Manual 1.04 Page 46 of 65...
  • Page 49 <value> FREE or RAMP or POINT Example :CHn:MOD:SWEEP:FREQ:TRIG:FREE Sweep Frequency Free Running Set :MOD:SWEEP:FREQ:TRIG? Query Synthesizer Sweep Frequency Trigger Example :CHn:MOD:SWEEP:FREQ:TRIG? FREQ SWEEP TRIGGER FREE <or> FREQ SWEEP TRIGGER RAMP <or> FREQ SWEEP TRIGGER POINT www.holzworth.com HS9000B Series Manual 1.04 Page 47 of 65...
  • Page 50 Set Synthesizer Sweep Frequency Direction <value> UP or DOWN Example :CHn:MOD:SWEEP:FREQ:DIR:UP FREQ SWEEP DIRECTION UP :MOD:SWEEP:FREQ:DIR? Query Synthesizer Sweep Frequency Direction Example :CHn:MOD:SWEEP:FREQ:DIR? FREQ SWEEP DIRECTION UP or FREQ SWEEP DIRECTION DOWN www.holzworth.com HS9000B Series Manual 1.04 Page 48 of 65...
  • Page 51 Query Synthesizer Sweep Dwell Time Example :CHn:MOD:SWEEP:FREQ:DWL? 700 us :MOD:SWEEP:FREQ:DWL:MAX? Query Synthesizer Maximum Sweep Dwell Time Example :CHn:MOD:SWEEP:FREQ:DWL:MAX? 10000000 us :MOD:SWEEP:FREQ:DWL:MIN? Query Synthesizer Minimum Sweep Dwell Time Example :CHn:MOD:SWEEP:FREQ:DWL:MIN? 100 us www.holzworth.com HS9000B Series Manual 1.04 Page 49 of 65...
  • Page 52 Set Synthesizer Sweep Number of Points <value> Synthesizer Dependent Example :CHn:MOD:SWEEP:FREQ:PTS:50 Sweep Frequency Points Set :MOD:SWEEP:FREQ:PTS? Query Synthesizer Sweep Points Example :CHn:MOD:SWEEP:FREQ:PTS? :MOD:SWEEP:FREQ:PTS:MAX? Query Synthesizer Maximum Sweep Points Example :CHn:MOD:SWEEP:FREQ:PTS:MAX? 65535 www.holzworth.com HS9000B Series Manual 1.04 Page 50 of 65...
  • Page 53 Set Wide Band List Number of Points <value> Synthesizer Dependent Example :CHn:MOD:LIST:WIDE:PTS:500 Wide Band Points Set :MOD:LIST:WIDE:PTS? Query Wide Band List Points Example :CHn:MOD:LIST:WIDE:PTS? :MOD:LIST:WIDE:PTS:MAX? Query Maximum Wide Band Points Example :CHn:MOD:LIST:WIDE:PTS:MAX? 3232 www.holzworth.com HS9000B Series Manual 1.04 Page 51 of 65...
  • Page 54 Set Wide Band Dwell Time command. The list of dwell times is not saved to the device. If the synthesizer is power cycled, then the complete list with dwell times must be reloaded. www.holzworth.com HS9000B Series Manual 1.04 Page 52 of 65...
  • Page 55 Example :CHn:MOD:MODE:LIST:WIDE:FREE Wide Band Free Running Set :MOD:MODE:LIST:WIDE? Query Wide Band Trigger Example :CHn:MOD:MODE:LIST:WIDE? WIDE LIST MODE TRIGGER FREE <or> WIDE LIST MODE TRIGGER LIST <or> WIDE LIST MODE TRIGGER POINT www.holzworth.com HS9000B Series Manual 1.04 Page 53 of 65...
  • Page 56 :CHn:MOD:LIST:WIDE:DWL:MIN? 100 us *NOTE: If a dwell time is loaded with each point in Set Wide Band List Values, then the value for Set Wide Band Dwell Time will be ignored. www.holzworth.com HS9000B Series Manual 1.04 Page 54 of 65...
  • Page 57 Set Narrow Band List Number of Points <value> Synthesizer Dependent Example :CHn:MOD:LIST:NARROWPTS:300 Narrow Band Points Set :MOD:LIST:NARROW:PTS? Query Narrow Band List Points Example :CHn:MOD:LIST:NARROW:PTS? :MOD:LIST:NARROW:PTS:MAX? Query Maximum Narrow Band Points Example :CHn:MOD:LIST:NARROW:PTS:MAX? 3232 www.holzworth.com HS9000B Series Manual 1.04 Page 55 of 65...
  • Page 58 Set Narrow Band Dwell Time command. The list of dwell times is not saved to the device. If the synthesizer is power cycled, then the complete list with dwell times must be reloaded. www.holzworth.com HS9000B Series Manual 1.04 Page 56 of 65...
  • Page 59 Example :CHn:MOD:MODE:LIST:NARROW:FREE Narrow Band Free Running Set :MOD:MODE:LIST:NARROW? Query Narrow Band Trigger Example :CHn:MOD:MODE:LIST:NARROW? NARROW LIST MODE TRIGGER FREE <or> NARROW LIST MODE TRIGGER LIST <or> NARROW LIST MODE TRIGGER POINT www.holzworth.com HS9000B Series Manual 1.04 Page 57 of 65...
  • Page 60 :CHn:MOD:LIST:NARROW:DWL:MIN? 6 us *NOTE: If a dwell time is loaded with each point in Set Narrow Band List Values, then the value for Set Narrow Band Dwell Time will be ignored. www.holzworth.com HS9000B Series Manual 1.04 Page 58 of 65...
  • Page 61 1 PLL Locked <or> 0 PLL Unlocked <or> 0 PLL Disabled, External 100MHz <or> 0 PLL Disabled, Internal 100MHz When using an external 10MHz reference, the response will be „1 PLL Locked‟ or „0 PLL Unlocked‟ www.holzworth.com HS9000B Series Manual 1.04 Page 59 of 65...
  • Page 62: Appendix B: Ethernet Configuration Commands

    A Description of the command here. <value> Defined here, if any, queries typically have no value Example Example ASCII sent in transmission Example ASCII received back, if a receive transmission is made www.holzworth.com HS9000B Series Manual 1.04 Page 60 of 65...
  • Page 63 TCP data communication can be established using the device IP address or the device host name. The host name is the complete serial number of the HS9000B synthesizer. An example host name is HS9005B-309. www.holzworth.com HS9000B Series Manual 1.04 Page 61 of 65...
  • Page 64 Query IP status Example :IP:STATUS? Static IP Address <or> DHCP :IP:ADDR:<value> Set Static IP Address <value> IP Address Example :IP:ADDR:192.168.10.11 Static IP address changed :IP:ADDR? Query Static IP Address Example :IP:ADDR? 192.168.10.11 www.holzworth.com HS9000B Series Manual 1.04 Page 62 of 65...
  • Page 65 Gateway address changed. :IP:GATEWAY? Query Gateway Address Example :IP:GATEWAY? 192.160.10.1 :IP:SUBNET:<value> Set Subnet for Static IP Address <value> IP Address Example :IP:SUBNET:255.255.0.0 Subnet address changed :IP:SUBNET? Query Subnet Address Example :IP:SUNET? 255.255.0.0 www.holzworth.com HS9000B Series Manual 1.04 Page 63 of 65...
  • Page 66: Appendix C: Gpib Configuration Commands

    <value> ON or OFF Example :GPIB:EOIWLC:OFF EOI with last character disabled :GPIB:EOIWLC? Query instrument GPIB EOI with last character Example :GPIB:EOIWLC? "EOI with last character enabled" OR "EOI with last character disabled" www.holzworth.com HS9000B Series Manual 1.04 Page 64 of 65...
  • Page 67 Set instrument GPIB response status <value> ON or OFF Example :GPIB:RESPOND:ON GPIB responds with every command :GPIB:RESPOND? Query instrument GPIB response status Example :GPIB:RESPOND? "GPIB only responds to queries" OR "GPIB responds with every command" www.holzworth.com HS9000B Series Manual 1.04 Page 65 of 65...
  • Page 68 All rights reserved. Holzworth Instrumentation Inc. 2540 Frontier Ave., STE 200 Boulder, Colorado 80301 USA +1.303.325.3473 www.HOLZWORTH.com 1.04 Version July 2020...

Table of Contents