5.4 ENVIRONMENTAL SPECIFICATIONS

Environmental specifications are based on component margins, thermal verification testing and
current draw tests. Production unit performance is not verified over temperature.
PARAMETER
Operating Temperature
Temperature Monitor Range
Channel Operating Temperature
AC Power Supply
Power Consumption
Chassis
Channel (per)
Warm-Up Time
1
Specifications are subject to change per the discretion of Holzworth Instrumentation, Inc
*NOTE: Amplitude accuracy may vary when operated outside of channel operating
temperature specified in section 6.3. Internal channel temperature should be monitored
using the ":TEMP?" command (see Appendix A). External cooling is recommended to
regulate internal temperature depending on ambient operating conditions and number of
integrated channels.
DESCRIPTION
Operating Environment
Humidity
Altitude
Vibration
Storage (Non-Operating)
Temperature
Humidity
Altitude
Vibration

6.0 HARDWARE/SOFTWARE INSTALLATION

This section outlines the basic installation requirements and procedures for the HSX
Series Multi-Channel Synthesizer application GUI and the hardware either via USB or
Ethernet.
First, the application GUI software must be installed. The GUI software is contained on
the thumb drive that was included with the synthesizer module. If the thumb drive was
lost another can be emailed or the software can be downloaded after contacting
Holzworth support via email at:
+1.303.325.3473 (option 2).
HSX Series User Manual 1.14
MIN
TYPICAL
0 C
-40 C
+35 C
+40 C
*
100 V
AC
5 W
17 W
10 min
SPECIFICATION (by design)
RH 20% to 80% at wet bulb temp. <29C (non-condensing)
0 to 2,000m (0 to 6,561 feet)
0.21 G-rms maximum, 5Hz to 500Hz
-10C to + 60C
RH 20% to 80% at wet bulb temp. <40C (non-condensing)
0 to 4,572m (0 to 15,000 feet)
Altitude
0.5 G-rms maximum, 5Hz to 500Hz
support@holzworth.com
www.holzworth.com
1
MAX
COMMENTS
+55 C
+85 C
+45 C
Amplitude accuracy valid range
240 V
50 – 60Hz
AC
20 min
20 C (ambient temp. dependent)