:MOD:LIST:NARROW:<point>,<freq><freq suffix>,[dwell
time][dwell suffix]
<point>
<freq>
<freq suffix>
[dwell time]
[dwell suffix]
Example
:MOD:LIST:NARROW?<point>
<point> Point location.
Example
TX:
RX:
*NOTE: If a dwell time is not specified with each point, then the value used for dwell time will be the
value set using the Set Narrow Band Dwell Time command.
The list of dwell times is not saved to the device. If the synthesizer is power cycled, then the
complete list with dwell times must be reloaded.
HS9000B Series Manual 1.04
Set Narrow Band List Values*
Point location.
:MOD:LIST:NARROW:PTS:
Synthesizer Dependent.
the first frequency point plus 5 percent.
GHz,MHz, kHz, Hz
Synthesizer Dependent OPTIONAL
ms, [us] OPTIONAL
TX:
:CHn:MOD:LIST:NARROW:2,996MHz,10us
RX:
Stored frequency and dwell time for point 2 <or> Invalid point
Query Narrow Band List Value (for the given point)
Cannot be greater than the value set using
:MOD:LIST:NARROW:PTS:
:CHn:MOD:LIST:NARROW?2
996.0000000 MHz,10us <or> Invalid point
www.holzworth.com
Set Narrow Band List Value (for the
given point)
Cannot be greater than the value set using
All frequency values must be less than
Page 56 of 65