7.3 TROUBLESHOOTING ETHERNET CONNECTIONS

Prior to proceeding below press CTRL+ALT+DEL to open Windows Task Manager. Click the
Processes tab. Ensure that there is only one instance of the application GUI open. If there are more
than one, end each Holzworth process, re-launch the GUI, and attempt to establish a connection.
7.3.1 Ethernet Reset via USB & Console Window
1. If the synthesizer is not discovered by the application GUI there may be static TCP/IP settings
that conflict with the current network configuration. If the static settings are unknown, USB
communication may be used to reset the synthesizer to DHCP or re-configure the static network
settings.
2. Establish a USB connection with the synthesizer as shown in section 8.1.
3. Launch the Console window using the button at the bottom right of the GUI. The Console can
be used to send ASCII commands to change static network settings or change from static mode to
DHCP and vice versa.
4. Refer to Appendix B for Ethernet configuration commands. Type commands into the text field
and then press Enter or click Send to send a command.
5. Begin by querying with the :IP:STATUS? command. Change status and/or re-configure the static
network settings as necessary.
6. Power cycle the HS9000B if prompted. Any status change from DHCP to Static or vice versa will
require a power cycle.