Keysight Technologies N9020B Installation Note page 46

Mxa signal analyzer
Table of Contents

Advertisement

Installation Procedure
6. Refer to
Figure
36. Place the rear panel (1) into position in the chassis. Replace the screws (2)
to attach the rear panel to the chassis. Torque to 9 inch-pounds.
Figure 36
Rear Panel Replacement
7. Refer to
Figure
37. Reattach the ribbon cable W1.
Figure 37
Front Panel Cable