Keysight Technologies N9020B Installation Note page 30

Mxa signal analyzer
Table of Contents

Advertisement

Installation Procedure
Installing New Coax Cables
Figure 19
Installing New Coax Cables
1. Refer to
Figure
19. Locate W48
between the input of the A12 Preselector and Port 4 of the SW4 Transfer Switch. Torque to 10
inch-pounds.
2. Locate W45
N9010-20009
Port 1 of the SW4 Transfer Switch. Torque to 10 inch-pounds
3. Locate W47
N9020-20129
the A12 Preselector and Port 3 of the SW4 Transfer Switch. Torque to 10 inch-pounds
4. Re-install W9 semi-rigid coax cable between the output of the A10 Input Attenuator B and
A1J1. Torque to 10 inch-pounds
5. Locate W46
N9020-20130
Port 2 of the SW4 Transfer Switch. Torque to 10 inch-pounds
6. Re-install W3 semi-rigid coax cable between A11J2 and A13J2. Torque to 10 inch-pounds.
30
N9020-20128
semi-rigid coax cable in the kit. Install W48
semi-rigid coax cable in the kit. Install W45 between A11J3 and
semi-rigid coax cable in the kit. Install W47 between the output of
semi-rigid coax cable in the kit. Install W46 between A13J9 and
Installation Note N9020-90289