Keysight Technologies N9020B Installation Note page 37

Mxa signal analyzer
Table of Contents

Advertisement

14.Refer to
Figure
27. Disconnect cable W12 from A16J710 on the A16 Reference Assembly and
discard.
Figure 27
Reference Cables
15.Refer to
Figure
27. Locate W54 in the Opt BUD Cable Kit. This cable has part number
8121-2288 and will be labeled "716" on one end and "14" on the other. Connect the end
labeled "716" to A16J716 on the A16 Reference Assembly. Route the cable along the right side
wall to A3J14 on the A3 DIF Assembly. Add this cable to the spiral wrap around W14.
16.Refer to
Figure 23
part number 8120-8868 and will be labeled "601" on one end and "19" on the other. Connect
the end labeled "19" to A3J19 on the A3 DIF Assembly. Route the cable to the right side of the
instrument and up to the A2 EAIF Assembly. Connect the cable to A2J601 with the cable
routed behind the screw head near A2J601.
17.Refer to
Figure
23. Connect W14 to A2J300 on the A2 EAIF Assembly.
Installation Note N9020-90289
Refer to
"Analyzer Information"
step 15
and
step 17
if the analyzer already has Options DP2 and/or MPB installed. Do perform
step 16 in all cases.
and
Figure
24. Locate W41 in the Opt BUD Cable Kit. This cable will have
at the beginning of the installation procedure. Do not perform
Installation Procedure
37