Installation Procedure
Rear Panel Removal
1. Refer to
Figure
8. Using the T-10 driver, remove the screws (2) attaching the rear panel (1) to
the chassis. The rear panel can now be removed.
2. Use a 9/16-inch socket wrench to remove the nut securing the EXT REF IN connector from the
rear panel.
Figure 8
Rear Panel Removal
3. Instruments with Option CR3 or CRP will have an additional cable W39 that is attached to the
rear panel. Remove W39 using either a 5/16" nut driver or a 5/16" open-end wrench.
4. The rear panel can now be removed.
20
Installation Note N9020-90289

Advertisement