Positioning the components in the load carrier

Depending on the equipment level of the device,
not all components may be present.
All components must be positioned securely in the
load carrier. Ensure the following:
– All surfaces and interiors can be completely
rinsed.
– The water can drain off freely.
– The hoses to be reprocessed are connected to
a flushing nozzle on one side.
The following illustrations show the connections for
the individual components. Always use the
connections indicated. However, other positions
are possible depending on the washer-disinfector
used.
Flexible breathing bag arm (option)
Required aids:
– Hose (outer diameter: 22 mm) for connecting
the breathing bag arm
C
B
A
1 Connect the flexible breathing bag arm (A) to
the hose (B).
2 Place both in the load carrier.
3 Connect the other end of the hose to a suitable
nozzle (C) on the load carrier.
Reprocessing Fabius family
Absorber insert
Required aids:
– Hose (outer diameter: 22 mm) for connecting
the absorber insert
A
B
C
1 Connect the absorber insert (A) to the hose (B).
2 Place both in the load carrier.
3 Connect the other end of the hose to a suitable
nozzle (C) on the load carrier.
Absorber container
 Place the absorber container with the opening
facing downwards on an injector nozzle.
English
21