Preparing the inspiratory valve and the
expiratory valve
A
B
C
D
E
Inspiratory valve:
1 Remove the valve cover plug (A) or the O
sensor from the dome (C) of the inspiratory
valve.
2 Screw off the union nut (B).
3 Remove the dome (C).
4 Take out the valve plate (D). The sealing ring
(E) remains on the valve crater.
5 Refit the dome (C) and screw on the union nut
(B).
Expiratory valve:
6 Screw off the union nut (F).
7 Remove the sight glass (G).
8 Take out the valve plate (H). The sealing ring (I)
remains on the valve crater.
9 Refit the sight glass (G) and screw on the union
nut (F).
14
Removing the APL valve
WARNING
Risk of damage to the compact breathing sys-
tem
F
If the APL valve is not disassembled before
the compact breathing system is reprocessed,
G
this can lead to leakages in the compact
breathing system.
H
Always remove the APL valve prior to repro-
I
cessing.
2
1 Loosen the knurled nut (A).
2 Remove the valve.
A
Reprocessing Fabius family