Fitting the anesthetic gas receiving system
The anesthetic gas scavenging system (AGSS)
must conform to ISO 8835-3 and ISO 80601-2-13.
WARNING
Risk of patient injury
If the side openings of the receiving system
are covered, this may cause a negative pres-
sure in the breathing system and in the pa-
tient's lungs.
Always make sure that the side openings on
the receiving system are not blocked.
A
C
B
 Observe the instructions for use of the
anesthetic gas receiving system (AGS).
1 Fit the receiving system on Fabius with the slots
resting on the corresponding pegs. Slide the
system down.
2 Use the screw plug (A) to seal the connection
not in use.
3 Slide the transfer hose (B) to the provided port.
4 Connect the other end of the transfer hose to
the exhaust port on the bottom side of the
compact breathing system.
5 Connect the suction hose (C) to the
corresponding port of the receiving system.
Reprocessing Fabius family