Dräger Fabius Tiro SW 3.n Reprocessing Manual page 19

Anesthesia workstation, software 3.n
Hide thumbs Also See for Fabius Tiro SW 3.n:
Table of Contents

Advertisement

Preliminary cleaning
Preliminary cleaning has been performed for
selected components according to the following
table.
Step
Component Agent
Preliminary
Compact
cleaning
breathing
system (pre-
pared)
Positioning the compact breathing system in
the ultrasonic bath
Prerequisite:
– The compact breathing system has been
prepared:
– All hoses and cables have been removed,
see p.10.
– The CO2 absorber has been removed, see
p.11.
– The CLIC adapter (option) has been
removed, see p.12.
– The flexible or rigid breathing bag arm
(option) has been removed, see p.12.
– The exhaust port has been removed, see
p.12.
– The flow sensor has been removed, see
p.13.
– The inspiratory valve and the expiratory
valve have been prepared, see p.14.
– The APL valve has been removed, see
p.14.
Reprocessing Fabius family
Manufactur-
er
Sekusept ak-
Ecolab
tiv
Concentra-
Tempera-
tion
ture
2 %
English
Contact
time
5 min in an
ultrasonic
bath at
35 kHz +
rinse for
5 min with
demineral-
ized water
19