Dräger Fabius Tiro SW 3.n Reprocessing Manual page 11

Anesthesia workstation, software 3.n
Hide thumbs Also See for Fabius Tiro SW 3.n:
Table of Contents

Advertisement

Removing the CO
absorber
2
1 Remove the CO
absorber:
2
– CLIC absorber (disposable absorber)
Or
– Reusable absorber
CLIC absorber (disposable absorber)
Observe the instructions for use of the CLIC
absorber.
C
1 Press the release button (A). The CLIC adapter
flips open.
2 Pull the CLIC absorber (B) upwards out of the
holder (C).
3 Flip the holder back until it audibly clicks into
place.
Reprocessing Fabius family
Reusable absorber
B
A
B
1 Unscrew the CO
compact breathing system.
2 Remove and dispose of the disposable dust
filter (B) (option).
3 Empty the used soda lime. Dispose of the soda
lime according to its instructions for use.
4 Remove the absorber insert (C) from the
absorber container. The sealing ring (D)
remains on the absorber insert.
A
absorber (A) from the
2
C
D
English
11