English
A
1 Place the compact breathing system (prepared)
with the absorber opening (A) facing upwards in
the ultrasonic bath.
2 Gently move the compact breathing system
back and forth in the ultrasonic bath for
5 minutes. Make sure that all trapped air
bubbles can escape from the interior.
3 Afterwards, rinse the compact breathing system
for 5 min with demineralized water.
Machine reprocessing
Step
Component Agent
Cleaning
Flushing
All semi-criti-
cal compo-
nents (see
Disinfecting
p.7)
Drying
20
Manufactur-
er
neodisher
Dr. Weigert
MediClean
forte
Demineral-
ized water
Concentra-
Tempera-
tion
ture
Min. 0.30 %
Min. 55 °C
(131 °F)
Tap water
temperature
Min. 90 °C
(194 °F)
Reprocessing Fabius family
Contact
time
Min. 10 min
Min. 1 min
5 min
Drying time
depends on
the load