Notations Used In The Instruction Manual; Important Precautions When Using Wga-910A And Optional Units - KYOWA WGA-910A Series Instruction Manual

For options bcd, d/a, rs-485, bcd and d/a
Hide thumbs Also See for WGA-910A Series:
Table of Contents

Advertisement

NOTATIONS USED IN THE INSTRUCTION MANUAL

The following notations are used in the Instruction Manual for convenience.
Names described on the panel surface
Names described on the panel surface are expressed in double quotation mark " ".
Informational notes
Certain notations are used as necessary to attract your attention to information that requires special care when
handing the product, and to information provided for reference purposes.
Examples of Notations
NOTE
Essential precautions required when handling the product.
MEMO
Reference items required when handling the product.

IMPORTANT PRECAUTIONS WHEN USING WGA-910A AND OPTIONAL UNITS

NOTE
The output signal (voltage/current) when the transducer cable is broken.
When the transducer cable is broken, the product may output the following output signal (voltage/current).
Therefore, take countermeasures to avoid the adverse effect, from the output signals, on the other measuring
instruments.
・D/A output and BCD D/A output
Voltage output: Approx. 11.2 V
Current output: Approx. 22.1 mA
The output signal (voltage/current) and unbalance load
・D/A output and BCD D/A output
Voltage output: The unbalance load should be 2k ohm or more.
Current output: The unbalance load should be 500 ohm or less.
The communication
・RS-485
RS-232C communication is not availavel.
6

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wga-910a-1Wga-910a-2Wga-910a-3Wga-910a-4Wga-910a-12

Table of Contents