5. TROUBLE SHOOTING

In the event that the product would not operate as expected or in a stable manner, you should trace the factor(s) that may
be causing such a condition before concluding that the system has failed.
Items to be checked and countermeasure against various phenomena are described in the following table.
If troubles are not solved despite countermeasures, contact Kyowa or our representatives.
If the product is damaged due to any causes other than as written in the Instruction Manual or if the user has disassembled
or remodeled the product, Kyowa Electronic Instruments Co., Ltd. may not be able to offer repair service.
For troubles and countermeasures relating to the main unit, see WGA-910A INSTRUCTION MANUAL.

5-1. OPERATION STOPS AND ACCEPTS NO INSTRUCTION AT ALL

The operation failed due to noise.
Turn OFF the power and turn ON again.
If the same phenomenon still generates, take appropriate countermeasures for removing the noise from operating
environment, power supply, etc.
Moreover, use the shielded BCD output cables, shielded D/A output cables, and shielded RS communication cables
and connect them to the GND terminal.

5-2. THE BCD OUTPUT IS STABLE

The indicating value and analog monitor output are stable, and the BCD output is also stable.
The input cable is not connected to the [INPUT] connector correctly or is breaking.
Check the channels.
The indicating value and analog monitor output are not stable, but the BCD output is stable.
The BCD output cable is not connected to the connector correctly or is breaking.
Repair the BCD output cable or replace the cable with a new one.

5-3. THE D/A OUTPUT IS STABLE

The indicating value and analog monitor output are stable, and the D/A output is also stable.
The input cable is not connected to the [INPUT] connector correctly or is breaking.
Check the channels.
The indicating value and analog monitor output are not stable, but the D/A output is stable.
The D/A output cable is not connected to the terminal block correctly or is breaking.
Repair the D/A output cable or replace the cable with a new one.
If all this fails to recover normal operation, contact KYOWA or our representatives.