KYOWA WGA-910A Series Instruction Manual page 17

For options bcd, d/a, rs-485, bcd and d/a
Hide thumbs Also See for WGA-910A Series:
Table of Contents

Advertisement

NOTE
The "Holding Input" does not hold the indicating values.
The BCD data output and binary data output do not output decimal points.
When the OUTPUT-prohibited signal is "L (ON)", the transistor outputs are all OFF.
To load the EOC data (negative logic), load the BCD data, POL (negative polarity) data, and OVER data
within T1/2 (msec) from the rising edge ("L (ON)"
"H (OFF)").
To load data after holding the BCD data, make sure the DATA HOLDING signal is "L (ON)." Then load
the data 5 msec or later. The BCD data does not vary when the DATA HOLDING signal is "L (ON)."
MEMO
While setting the BCD data output and binary data output, the BCD data output and binary data output
hold the values immediately before the setting.
To load the multiple BCD data outputs connected in parallel successively, control the
OUTPUT-prohibited signal.
The BCD data output COM and binary data output COM are connected to the control input COM of the
WGA-910A.
When the indicating value is in the detecting block or holding block, as the WGA-910A outputs the
BCD data or binary data, the "Detecting Block Output" or "Holding Block Output" becomes ON.