Fine Adjustment Of The Force Balance - Leica RM CoolClamp Instructions For Use Manual

Electronically cooled universal cassette clamp
Table of Contents

Advertisement

Fig. 6
There are two versions of the specimen holder fixture, one with and one without specimen orientation.
The two versions can be interchanged.
The specimen orientation allows for simple position correction of the specimen surface when the
specimen is clamped into place.
To do so, proceed as follows:
1. Move the specimen holder fixture
handwheel
2. To release the clamping system, turn the screw
Allen key.
3. Push the guide
holder fixture
4. To clamp the specimen clamp, turn the screw
5.4

Fine Adjustment of the Force Balance

Tip
Installing the RM CoolClamp necessitates performance of a force balance; for additional information,
refer to the Instructions for Use of the respective rotary microtomes.
RM CoolClamp
1
(→ "Fig. 6-1")
(→ "Fig. 6-2")
and engage the handwheel lock.
(→ "Fig. 6-5")
of the RM CoolClamp
(→ "Fig. 6-1")
as far as it will go.
4
2
3
5
to the upper end position by turning the
(→ "Fig. 6-3")
counterclockwise using a size 4
(→ "Fig. 6-4")
from the left into the specimen
(→ "Fig. 6-3")
clockwise as far as it will go.
5
Operation
1
3
13