3
Installing the G1330B Thermostat
Installing the G1330B Thermostat
Stage 6: Turning on the Thermostatted Autosampler
1 Depress the power switch to turn on the two modules of the autosampler.
The power switch stays depressed (1) and a green indicator lamp in the power switch is on
N O T E
when the autosampler is turned on. When the line power switch stands out (Ø) and the
green light is off, the autosampler is turned off.
Stage 7: Update of Control Module Firmware
If the control module has a firmware revision A.01.30 or higher, you do not
need to update the control module firmware.
If you have a control module version A.01.30 or lower, update the firmware as
described.
1 Disconnect the control module before inserting the PC card.
2 Insert the PC card into the card slot of the control module.
3 Reconnect the control module for restarting it.
4 Press "System" (F5) - "Records" (F4). Highlight the LC-System line in the
5 Press "FW-Update" (F5).
6 Select the file for the firmware update (LCB202en.BIN).
7 Press "Execute" and select "Yes" to confirm loading of the new firmware.
8 Check that the correct firmware was loaded by pressing "System" (F5) -
9 Disconnect the control module and remove the PC card by pressing the
38
display using the up-down arrows.
The control module reboots and loads the firmware indicated by (.) and (*)
on the display. When finished with the update the control module reboots
again.
"Records" (F4).
card-eject button.
1200 Series AT User Manual