Installing the G1330B Thermostat
Installing the G1330B Thermostat
Stage 1: Preparing the Autosampler Thermostat and Autosampler
Damage through condensation
WA R N I N G
If the condensation tube is located in liquid the condensed water cannot flow out of
the tube and the outlet is blocked. Any further condensation will then remain in the
instrument. This may damage the instruments electronics.
➔
Even under average humidity conditions, a significant amount of condensed water gathers
N O T E
every day. A suitable container must be provided and emptied regularly in order to avoid
overflow.
1 Place the autosampler thermostat on the bench or in the stack.
2 Remove the front cover. Press the two snap fasteners on the sides of the
26
Make sure that the condensation tube is always above the liquid level in the vessel.
cover and move it away.
1200 Series AT User Manual