Operation; Before Operation; Turning On The Nitrogen Generator - Agilent Technologies 4107 User Manual

Nitrogen generator
Table of Contents

Advertisement

Before Operation

Before starting the nitrogen generator, carefully read the 'Safety
Practices and Hazards' section at the front of this guide and ensure
that the laboratory is set up according to the details specified in the
Agilent MP-AES Site Preparation Guide.
In operation, the nitrogen generator cycles approximately every 40
seconds. At the end of each cycle the oxygen enriched air is vented to
the room from a vent in the bottom of the nitrogen generator. A noise
may be heard as this gas is released.

Turning on the Nitrogen Generator

To turn on the nitrogen generator:
1
2
3
Agilent 4107 Nitrogen Generator User's Guide

4. Operation

Before Operation
Turning on the Nitrogen Generator
Selecting the Nitrogen Generator
Purging the Nitrogen Generator
Turning Off the Nitrogen Generator
Depressurizing the Nitrogen Generator
Ensure that the gas filters are installed correctly and do not need
replacing.
Ensure the air and nitrogen hoses are connected to the nitrogen
generator and that the air supply is turned on.
Check the pressure indicated on the pressure gauge of the filter
regulator unit on the right side of the nitrogen generator meets
the specification.
Operation
21
21
22
23
23
24
21