Condensation drain tube
Waste bottle
Figure 7
3 Install the LAN interface board in the sampler (if required), see
4 Remove the adhesive tape which covers the front door.
5 Remove the front door and remove the transport protection foam.
6 Place the Autosampler on top of the Thermostat. Make sure that the
7 Place the air channel adapter into the autosampler tray base. Make sure the
Agilent 1200 Series Autosampler Reference Manual
Condensation Leak outlet
Communication Interface Board"
Autosampler is correctly engaged in the Thermostat locks.
adapter is fully pressed down. This assures that the cold airstream from the
Thermostat is correctly guided to the tray area of the Autosampler.
Installing the Autosampler
on page 245.