Download Print this page

Hach ORBISPHERE 3654 Basic User Manual page 11

Portable analyzer
Hide thumbs Also See for ORBISPHERE 3654:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
4.8 Flow chamber installation
At the top of the TC sensor is a thermistor to measure the sample temperature. Do not bend or damage this
component.
The model 32013A or 32015 flow chamber allows the liquid or gaseous sample past the TC and
temperature sensors. The flow chamber attaches to the TC sensor with a threaded collar and is
sealed to the sensor with two O-rings (EPDM Ø20 × 2 mm and NBR Ø 4 × 1.5 mm).
The flow chamber's inlet and outlet use ¼-inch diameter transparent plastic tubing. Connect by
compression fittings to the sample source and to the drain, respectively.
You may also have received a model 32051 sample tube adapter to attach the flow chamber inlet
tubing to the sampling point.
Figure 4 Flow chamber 32013A
1 Purge gas cylinder refill protection cap
2 Sensor locking pin
3 TC sensor
4 Flow rate adjustment valve
4.9 Instrument location
Locate the instrument close to the sample being analyzed, and to the PC if one is being used. A
LEMO-6 connector for RS-232 serial output to a PC is on the right side of the instrument. Refer to
Figure 1
on page 8.
Make sure you install a fully charged set of batteries, or connect the instrument to an external power
source before switching the instrument on and exposing the sensor's membrane to any liquid. Refer
to
External power supply (optional)
C A U T I O N
5 Outlet
6 Flow chamber
7 Flow chamber collar
8 Purge gas exit tube
on page 9.
English 11

Hide quick links:

Advertisement

loading