Setting Up the 86122A
N O TE
28
2 Touch the center of the X with your finger or a stylus. The X will move to the
opposite corner of the display screen. When the display is calibrated, the touch
screen configuration utility closes.
The calibration will time out within 10 seconds if you do not touch the
center of the X. You can also press
touch screen calibration.
If you miss the center of the
display an error message and revert back to the previous calibration. This
will prevent the touch screen from being rendered useless due to a poor
calibration.
General Safety Considerations
on an optional keyboard to exit the
Esc
by a significant amount, the instrument will
X
Agilent 86122A Multi-Wavelength Meter, Second Edition