Agilent 86122A Multi-Wavelength Meter, Second Edition
The 86122A Multi-Wavelength Meter uses a specially designed
Windows 98 application program (Windows ® is a U.S. registered
trademark of Microsoft Corp.) All 86122A Multi-Wavelength Meter
functionality is directly available from within the 86122A Multi-
Wavelength Meter application. Windows 98 configuration changes
made outside of the 86122A Multi-Wavelength Meter application
may not work correctly and could cause the instrument to become
inoperable. Do not try to access or make changes to the Windows
Operating system. Repairs caused by the improper use of the
86122A Multi-Wavelength Meter will not be covered under
warranty.
VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be
restricted. The ambient temperature (outside the cabinet) must be
less than the maximum operating temperature of the product by 4° C
for every 100 watts dissipated in the cabinet. If the total power
dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
Install the instrument so that the detachable power cord is readily
identifiable and is easily reached by the operator. The detachable
power cord is the instrument disconnecting device. It disconnects
the mains circuit from the mains supply before other parts of the
instrument. The front panel switch is only a standby switch and is
not a LINE switch. Alternatively, an externally installed switch or
circuit breaker (which is readily identifiable and is easily reached by
the operator) may be used as a disconnecting device.