Agilent Technologies 86122A User Manual page 105

Multi-wavelength meter
Table of Contents

Advertisement

Specifications
Table 9
Description
[1]
Selectivity
Input Power
Maximum Displayed Level
Maximum Safe Input Level
Return Loss
With Straight Connectors (PC, Opt. 021)
With Angled Connectors (APC, Opt. 022)
Measurement Cycle Time
Maximum Number of Lines
Measurement Modes
Delta Modes
Built in Automatic Measurement Applications
[1][4]
Signal to Noise Ratio
Channel Spacing
100 GHz
Channel Spacing
50 GHz
Drift
Fabry-Perot Characterization
Additional Features
Inputs/Outputs
Optical Input
Rear Panel Connectors
Agilent 86122A Multi-Wavelength Meter, Second Edition
86122A
25 dB spacing
10 dB spacing
+10 dBm
+18 dBm
35 dB
50 dB
0.5 s
1000
List by wavelength table, list by power table,
peak wavelength and power, average
wavelength and total power
Delta wavelength, delta power, delta
wavelength and power
>35 dB with 100 averages
>27 dB with 100 averages
Maximum, minimum, total drift (max-min) of
wavelengths and powers over time
Mean wavelength, peak wavelength, mode
spacing full-width half maximum, peak
amplitude total power, sigma
Power offset, power bars (on or off), user
adjustable peak excursion and peak
threshold, user adjustable start and stop
wavelength limits, graphical display, save and
recall instrument states.
9/125 µm single-mode fiber
LAN, PS/2 for Keyboard & Mouse, SVGA for
external monitor, GPIB, parallel printer port,
AC Line, optional optical input
Specifications and Regulatory Information
90 GHz
10 GHz
Sum of all lines input
Number of laser lines may
be limited by signal power
requirements for accurate
wavelength measurements.
Data Logging and sorting
by any parameter are
included in the 86122A.
0.1 nm noise bandwidth.
Lines above –25 dBm.
Notes