Examples Of Beta Response; Fast Neutron Response (Pube); Figure 1-2: Nominal Photon Energy Response - Thermo Scientific RO-20 Operating Manual

Aa ion chamber
Table of Contents

Advertisement

In the following examples, the RO-20 had been calibrated to
slide facing the source.
• ±30% from 8 keV to 1.3 MeV, with the open slide facing the source.
• ±15% from 33 keV to 1.3 MeV, with the closed slide facing the source.
• ±15% from 55 keV to 1.3 MeV, through the side of the instrument.
1.3.8

Examples of Beta Response

Uranium slab: 30% of true mrad/h field behind 7 mg/cm
resting on the slab (slide open).
90
90
Sr–
Y: Approximately 93% of true mrad/h field at 30 cm with slide open.
1.3.9

Fast Neutron Response (PuBe)

Reads approximately 8% in mR/h of true neutron field in mrem/h.

Figure 1-2: Nominal photon energy response

Ro-20
Löw 08.10.2008
137
Cs (slide closed), with the
RB-XXX-XXXXXX E
2
window, with the RO-20
1-3