Omron F3SG-SR Series Manual page 98

Safety light curtain
Hide thumbs Also See for F3SG-SR Series:
Table of Contents

Advertisement

F3SG-SR/PG
Description
Operating range selection setting
error
External device monitoring error
Error other than those above
*1. You can check the error codes by SD Manager 3 or SD Manager 3 Mobile APP.
*2. Other indicators than the ERR and MAINT indicators are not illuminated. For details of the error, refer to [Code] and [Error description] displayed
in [Error Log] in the SD Manager 3.
98
Checking by
SD Manager 3/
Indicator
SD Manager 3
Mobile APP
X
X
EB
---
X
52
2
*
Error code
---
X
other than
*
2
those above
Error code
(HEX)
*
1
The setting of the operating range selection may be
incorrect.
• When the Intelligent Tap is connected, check if the
Operating Range Selection of the DIP Switch is
properly set.
• When the Intelligent Tap is not connected, check if the
Operating Range Select Input line is properly wired.
Relay may be welded. Replace the relay.
The relay and the RESET line may not be properly wired.
Check the wiring with the relay.
The relay response time may be exceeding the allowable
delay time. Change the allowable delay time or replace the
relay with one that has an appropriate response time.
An error may have occurred in the internal circuit.
Replace the F3SG-SR.
Cause and measures

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

F3sg-pg seriesF3sg-4sr seriesF3sg-4pga seriesF3sg-4sra0160-14F3sg-4srb0160-14F3sg-4sra0200-14-f ... Show all

Table of Contents